Uncertainty calculations in optical methods used for micro flow measurement

Utilize este identificador para referenciar este registo.
Nome:Descrição:Tamanho:Formato: 
1_s2.0_S2665917421001185_main.pdf3.13 MBAdobe PDF Ver/Abrir

Orientador(es)

Resumo(s)

This work aims to describe the uncertainty calculation methodologies associated with two optical methods used for micro flow measurements, both developed at the Portuguese Institute for Quality (IPQ) under the EMPIR project MeDD II – Metrology for Drug Delivery framework. One of the methods is the front track which consists of tracking the meniscus of a liquid inside a capillary tube over time. The second method is the pending drop method that relies on measuring the volume growth of a drop over time. Both methods use a camera to capture images and convert them in volume or displacement length to determine flow rate. The uncertainty calculations will be presented in detail. The values obtained for the front track method are much smaller than for the pending drop method, specially at very low flow rates.

Descrição

Funding Information: This work performed under MeDDII project has received funding from the EMPIR programme co-financed by the Participating States and from the European Union's Horizon 2020 research and innovation programme .

Palavras-chave

Front tracking Measurement uncertainty Method Microflow Pending drop method Electrical and Electronic Engineering Industrial and Manufacturing Engineering Mechanics of Materials Electronic, Optical and Magnetic Materials